نفت و پتروشیمی

دستگاه تعیین نقطه قطره ای شدن گریس
- درجه حرارت: از دمای محیط به 290 درجه سانتی گراد (554 درجه فارنهایت)
- منبع تغذیه: ۲۲۰ ولت و ۵۰/۶۰ هرتز

- قدرت: 700 وات
- ابعاد: ۱۸*۲۱*۵۲سانتیمتر
- وزن: 3 کیلوگرم

ساخت کمپانی SDM ایتالیا

Dropping Point

November 30th, -0001

For determination of the dropping point of lubricating grease. This test method is not recommended for use at bath temperatures above 288°C (550°F). For higher temperatures Test Method D 2265 should be used.

This specification covers lubricating greases suitable for the periodic relubrication of chassis systems and wheel bearings of passenger cars, trucks, and other vehicles.

Consists of a nickel-plated brass casing, test tube with projection cap and cork ring, thermometer depth gage, cup plug gage, cover of anticorodal, a polished metal rod Ø1.5x150mm dia., 400ml beaker, heating device unit with motor stirrer.

- Temperature: from ambient to 290°C (554°F)
- Power supply: 220V 50/60Hz
- Power: 700W
- Dimensions: 18x21x52 cm
- Weight: 3 kg

Drop Melting Point

June 17th, 2017

Drop Melting Point of Petroleum Wax Including Petrolatum.

This test method covers the determination of the drop melting point of petroleum wax. It is used primarily for petrolatums and other microcrystalline wax.

Art. LT/DM-210000/M

Drop Melting Point ASTM D127

  • Heating device with a stainless steel housing and with electronic regulator
  • 1 transparent container (jar) of 1500 ml capacity supplied with Teflon cover for the accommodation of the test tube by means of a cork stopper having a hole in the center for introducing and holding the test tube
  • Teflon cover is also equipped with a filling hole with cap and hole with relevant metal holder for thermometer
  • Test tube with 25 mm outside diameter and 150 mm long with cork stopper with hole for thermometer

Power supply

  • 230 Vac 50/60 Hz
  • Power consumption 1000 Watt

Accessories

  • T-AS14C: thermometer ASTM 14C
  • T-AS34C: thermometer ASTM 34C
  • T-AS61C: thermometer ASTM 31C

Spare Parts

  • LAB-102-101: test tube, pack of 10 pcs.
  • LAB-102-102. Pyrex® Jar
  • LAB-102-103: cork stopper series
  • LAB-102-104: Teflon cover

The dropping point is the temperature at which the grease passes from a semisolid to a liquid state under the conditions of test. Cooperative testing indicates that in general, dropping points by test method D 566 is in agreement. The test, usually carried on with manually operated analyzers, has been totally automated with this instrument: in fact it provides automatically to regulate the heating rate, beginning with a 5.5°C/min rate and then reducing it to 1°C/min rate.
The apparatus is a benchtop model which houses the components and a Panel PC with touch screen. A software running under Microsoft Windows XP permits to select the test method and the test parameters, run the test automatically, store, retrieve and print data, diagnose and calibrate the instrument offering in the meanwhile all the features of Windows systems such as LAN connectivity. A unique cooling system permits to cool down quickly the heating furnace allowing to start a new test in a few minutes.